Eddy current on turbine blades finds cracks only if the probe, frequency, and standard still match
← All analysis

Sector · Energy · 03 Sept 2026 · Brief

Eddy current on turbine blades finds cracks only if the probe, frequency, and standard still match

Lift-off, worn shoes, and skipped calibration shift the impedance plane before the first false clear. A green “scan complete” does not prove the blade root was interrogated.

Outage NDE packages still close with “ECT OK” while the instrument was never zeroed on the current shoe. Frequency was left from the last alloy family. A subsurface indication at the fir-tree root waited for the next start.

Eddy current owns coil coupling and calibrated response, not waving a probe. Geometry, lift-off, and reference standards decide whether a crack is a blip or invisible.

Gate before you clear a row

| Item | Why | Fail action | |------|-----|-------------| | Same shoe / frequency as procedure | Wrong setup misses depth | Reconfigure; re-run standards | | Calibration on notch standards | Drift hides small cracks | Recal; do not proceed | | Lift-off check on curved airfoil | Gap kills sensitivity | Reseat; use correct shoe | | Technician cert + blade map | Unmapped skips are silent misses | Complete map before sign-off |

Eddy current impedance plane with calibration standards for blade work

Discipline

Log probe ID, frequency, and standard set on every blade row. Do not carry yesterday’s setup into a different alloy or coating without requalifying. Coating thickness changes are MoC for ECT—not a footnote.

The probe moves. The calibration decides whether that motion was an inspection.